Service Manual Service Manual SDS1000D Digital Storage Oscilloscope June 2011 SIGLENT Technologies Co., Ltd.

Página 2

Service Manual Hard Ware Frequency Counter Reading resolution 6 Bytes Range DC Couple, 10Hz to MAX Bandwidth Sign

Página 4

Service Manual Chapter 2 Prepare for Use This chapter includes the following contents: ● General Inspecting ● Power-On Inspecting ● Connect the

Página 6 - Catalogue

Service Manual Power-On Inspection Normal operating voltage for DS1000CA series digital oscilloscope is the range of 100-240VRMS with the frequency f

Página 8 - )1% of offset +0.2div+100mv]

Service Manual Auto Setup SDS1000D Series Digital Storage Oscilloscopes have auto setup function, which can identify the type of waveform and adjust

Página 10

Service Manual Chapter 3 Performance Test This chapter includes the following contents: Interface Test USB Host USB Device RS-232

Página 13 - General Inspecting

Service Manual USB Device Test Purpose: Test if the USB Device interface works normally through EasyScope for SDS1000D series oscilloscope. Tool

Página 15 - Probe Compensation

Service Manual Picture 3-6 EasyScope connect setting 4. Press “UTILITY” page3/4 IO Setting, select the Baud as 38400 5. Click “Connect”

Página 17 - Service Manual

Service Manual Specification Test Purpose: Test if some specifications are in the specified range. Table 3-2 require instruments Description Minim

Página 19 - Interface Test

Service Manual Test Trigger Sensitivity Steps:::: 1、Press the button “DEFAULT/SETUP”, and set the VOLT/DIV to 500mv/div, set SEC/DIV to

Página 21 - RS-232 Test

Service Manual Picture 4-2 Calibrating 4. After the calibration finished, Press “SINGLE” key to exit. Picture 4-3 Calibration succeed interface

Página 24 - Test DC Gain Accuracy

Service Manual Notice Notices:::: ● Don’t disassemble the products except the work need. ● Disassemble only can be done by qualified person.

Página 26 - Test Noise

Service Manual Disassemble and Assemble Back Rind Picture 5-2 Disassemble and Assemble back rind 1 图 5-3 Disassemble and Assemble back rind 2 D

Página 28

Service Manual Disassemble and Assemble Top Cover Board Picture 5-5 Disassemble and Assemble top cover board Disassemble steps: 1. Back out the

Página 30

Service Manual Disassemble and Assemble power board Picture 5-7 Disassemble and Assemble power board Disassemble steps: 1. Back out the four bo

Página 32

Service Manual Disassemble and Assemble Main board Picture 5-9 Disassemble and Assemble main board Disassemble steps: 1. Back out the five bolts

Página 34

Service Manual ●●●● Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present. ●●●● Do Not Operate With Suspec

Página 35 - Disassemble and Assemble Fan

Service Manual Disassemble and Assemble Keyboard Picture 5-11 Disassemble and Assemble keyboard Disassemble steps: 1. Back out the four bolts ○

Página 37

Service Manual Drive. Store Data Success! : Save setup data, waveform data or picture data to the internal of the oscilloscope or USB flash succe

Página 39 - Disassemble and Assemble LCD

Service Manual Chapter 7 Updating Parts SIGLENT Provide some replaceable parts in order to maintain or update for users. Please see as the followin

Página 41 - Prompting messages:

Service Manual Catalogue Document Summarize...

Página 43 - Troubleshooting

Service Manual Horizontal System Real Time Sampling Rate Single Channel:500MSa/s,Double Channel: 250MSa/s Equivalent Sampling Rate 50GSa/s Measu