Service Manual
Specification Test
Purpose: Test if some specifications are in the specified range.
Table 3-2 require instruments
Description Minimum requirements Examples
DC Voltage Source
-6 mV to 30 V, ±0.1% accuracy
Standard sine wave
generator
10Hz and 300MHz, ±1% amplitude
accuracy
FLUKE 9500B
50 Ω BNC
Feed-Through
Termination
DC to 1GHz PL-50
Before beginning these procedures, two conditions must first be met:
The oscilloscope must have been operating continuously for thirty minutes within the
operating temperature range specified.
You must do the Self Calibration procedure. If the ambient temperature changes by more
than 5℃, you must do the Self Calibration procedure again.
Picture 3-9 Test instrument connecting
Test DC Gain Accuracy
Steps 1Set the DC voltage source output level to 30 V
2Press the button “DEFAULT/SETUP” of oscilloscope
3Set VOLT/DIV to 10V/div, set measure parameter as “CH1,Mean”
4Connect the oscilloscope channel 1 to the DC voltage source, and record the value of
Mean appeared on the screen as V
pos
.
5Change the polarity of the DC voltage sourceand record the value of Mean appeared
on the screen as V
neg
.
6Calculate V
diff
=V
pos
V
neg
then compare V
diff
to accuracy limits in the table.