SIGLENT SDS1000D Manual de Serviço Página 20

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 19
Service Manual
USB Device Test
Purpose: Test if the USB Device interface works normally through EasyScope for SDS1000D
series oscilloscope.
Tools
An Unit SDS1000D digital oscilloscope
A computer with USB interface
A standard USB cable(Type AB)
EasyScope software for SDS1000D series
Steps
1Install the EasyScope to computer.
2Connect the oscilloscope with PC using an USB cable and install the driver program step by
step following the instruction.
Picture 3-3 connecting the USB Device interface
3Run EasyScope ,then click “Connect”, if successful, the two point lamps at the top right
corner of the software will be changed from red to green.; if failed, a prompt message will be
appeared as below.
Picture 3-4 USB Device connect failed
Vista de página 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 43 44

Comentários a estes Manuais

Sem comentários