40
display updates from left to right. There is no trigger or horizontal position control of
waveforms during scan mode.
2.7.2 Window Zone
Use the Window Zone option to define a segment of a waveform to see more detail. The
Window time base setting cannot be set slower than the Main time base setting.
You can turn the Horizontal Position and SEC/DIV controls to enlarge or minish
waveforms in the Window Zone.
“M” Mean main time base, “W” mean window time base.. It is also a arrow on the scale
top to show the vertical position.
If you want to see a section of the waveform in detail, follow these steps:
SDS1000C/CE/CF/CFL series Operate steps:
1. Press the “HORI MENU” button to enter the “Horizontal menu”.
2. Press the “Win Zone” option button.
3.
Turn the “S/div” knob (adjust window’s size) and turn the “Horizontal Position”
knob (adjust window’s position) to select the window that your need (see picture
2-27). The Window time base setting cannot be set slower than the Main time
base setting.
Picture 2.7-2