SIGLENT SDS1000D Manual do Utilizador Página 53

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 52
45
2. Set up Source
According to input signal, press the “Source” option button to select “CH1”,
“CH2”, “CH3”, “CH4”,“EXT”, “EXT/5”or “AC Line”.
3. Set up Slope
Press the “Slope” option button to select , “ ” or ”.
4. Set up Trigger mode
Press the “Trigger mode” option button to select “Auto”, “Normal”,
“Single”.
Auto: The waveform refresh at a high speed whether the trigger condition
is
satisfied or not.
Normal: The waveform refresh when the trigger condition is satisfied and
waits for next trigger event occurring when the trigger condition is
not satisfied.
Single: The oscilloscope acquire a waveform when the trigger condition is
satisfied and then stops.
5. Set up Trigger coupling
a. Press the “Set up” button to enter the “Trigger Setup Menu”.
b. Press the “Coupling” option button to select “DC”, “AC”, “HF Reject”
or “LF Reject”.
Pulse TriggerUse Pulse Width triggering to trigger on aberrant pulses.
Table 2-15 Pulse Trigger function Menu 1
Option Setting Explain
Type Pulse Select the pulse to
trigger the pulse match
the trigger condition.
Source CH1
CH2
CH3
CH4
EXT
EXT/5
Select input signal
source.
Vista de página 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 153 154

Comentários a estes Manuais

Sem comentários